Low Flow MicroBlenderandHigh Flow MicroBlenderInstruction Manual
MicroBlender Instruction Manual6 L2965–101 Rev. DHigh Flow MicroBlender
Instruction Manual MicroBlenderL2965–101 Rev. D 7Section 3: Warnings, Cautions, and NotesThe MicroBlender should be operated by trained, qualified med
MicroBlender Instruction Manual8 L2965–101 Rev. DWarning! If either the air or oxygen gas source fails, the MicroBlenderalarm sounds, alerting the cl
Instruction Manual MicroBlenderL2965–101 Rev. D 9Section 4: Controls and AlarmsThe MicroBlender delivers selected oxygen concentrations through two ou
MicroBlender Instruction Manual10 L2965–101 Rev. DSection 5: Performance ChecksBefore placing the MicroBlender into clinical use, perform the followin
Instruction Manual MicroBlenderL2965–101 Rev. D 11High Flow MicroBlender Alarm / Bypass CheckAdjustment Response1. Connect the 50 ±5 PSIG* air/oxygen
MicroBlender Instruction Manual12 L2965–101 Rev. DReverse Flow Check1. Connect both gas supply hoses to the inlet connectors.2. Connect the oxygen hos
Instruction Manual MicroBlenderL2965–101 Rev. D 13Section 6: Troubleshooting GuideProblem Potential Cause Corrective ActionOxygenconcentrationdiscrepa
MicroBlender Instruction Manual14 L2965–101 Rev. DProblem Potential Cause Corrective ActionMicroBlender isaccurate onlywhen inlet gaspressures areequa
Instruction Manual MicroBlenderL2965–101 Rev. D 15Section 7: Cleaning and SterilizingNote:User is to consult with the manufacturer of the ETO equipmen
MicroBlender Instruction Manualii L2965–101 Rev. DThis document is protected by United States and International Copyright laws.This document may not b
MicroBlender Instruction Manual16 L2965–101 Rev. DSection 8: Maintenance and ServiceCaution!The MicroBlender should only be serviced or calibrated by
Instruction Manual MicroBlenderL2965–101 Rev. D 17Section 9: Parts and AccessoriesParts and AccessoriesMicroBlender BracketsPart No. Description04322
MicroBlender Instruction Manual18 L2965–101 Rev. DSection 10: Explanation of AbbreviationsAir/O2Mixture of Compressed Air and Oxygen°C Degrees Centigr
Instruction Manual MicroBlenderL2965–101 Rev. D 19Section 11: Specifications
MicroBlender Instruction Manual20 L2965–101 Rev. DGas Inlet Supply Pressure Low Flow 30 – 75 psigHigh Flow 30 – 75 psig. Output flowrate will be dimin
Instruction Manual MicroBlenderL2965–101 Rev. D 21Alarm Sound Generator Low Flow Reed AlarmHigh FlowAlarm Sound Intensity Low Flow 80 dB minimum at 1
MicroBlender Instruction Manual22 L2965–101 Rev. DSection 12: WarrantyTHE PRODUCTS OF CAREFUSION CORPORATION (CAREFUSION HEREIN) ARE WARRANTEDTO BE FR
Low Flow MicroBlenderandHigh Flow MicroBlender
Instruction Manual MicroBlenderL2965–101 Rev. D iiiContentsSection 1: Introduction...
MicroBlender Instruction Manualiv L2965–101 Rev. DLow Flow MicroBlenderandHigh Flow MicroBlender
Instruction Manual MicroBlenderL2965–101 Rev. D 1Section 1: IntroductionThe MicroBlender is a lightweight, compact, air-oxygen blender that providespr
MicroBlender Instruction Manual2 L2965–101 Rev. DSection 2: Operation OverviewThe MicroBlender provides selection of oxygen concentrations by means of
Instruction Manual MicroBlenderL2965–101 Rev. D 3Gas InletsThe ports located on either side of the unit allow flexibility for the clinician. TheMicroB
MicroBlender Instruction Manual4 L2965–101 Rev. DProportioning ModuleFrom the balance module, the gases flow into the proportioning module andmix acco
Instruction Manual MicroBlenderL2965–101 Rev. D 5Similarly, if the blender is set to deliver 100% concentration and AIR sourcepressure is reduced or l